PROVA Model 2000 SPECIFICATIONS: (23°C ± 5°C)
DC Current
Range Resolution Accuracy (of rdg) Overload Protection
400A 0.1A ± 1.5% ± 3dgts 3000 amps DC
0-2000A 1A ± 1.5% ± 3dgts 3000 amps DC
2000-2500A 1A ± 2.0% ± 3dgts 3000 amps DC
AC Current: (True RMS, Crest factor < 4)
Accuracy (of rdg) Overload Protection
Range Resolution 50/60 Hz 40Hz 1kHz
400A 0.1A ± 1.5% ± 5dgts ± 2.0% ± 5dgts 3000 amps AC
0-1000A 1A ± 2.0% ± 5dgts ± 2.5% ± 5dgts 3000 amps AC
1000-2100A 1A ± 2.5% ± 5dgts ± 3.0% ± 5dgts 3000 amps AC
Analog Output Current:
Range Output Accuracy Overload Protection
0-400A 1mV/A ± 2.5% ± 0.5A 600 volts AC
400A-2100A 1mV/A ± 2.5% ± 5A 600 volts AC
DC Voltage: (Autorange & Manual. Overload protection 800VAC for all ranges)
Range Resolution Accuracy (of rdg) Input Impedance
40mV 0.1 mV ± 1.5% ± 3dgts 10M?
4V 1 mV ± 1.5% ± 3dgts 5M?
40V 10 mV ± 1.5% ± 3dgts 5M?
400V 100 mV ± 1.5% ± 3dgts 5M?
600V 1V ± 1.5% ± 3dgts 5M?
AC Voltage: (True RMS, Crest Factor < 4. Input impedance 5 M?)
Accuracy Overload Protection
Range Resolution 50/60 Hz 40Hz 1kHz
400mV 0.1mV 1 1 1
4V 1 mV ± 1.5% ± 5dgts ± 2.0% ± 5dgts 800 volts AC
40V 10 mV ± 1.5% ± 5dgts ± 2.0% ± 5dgts 800 volts AC
400V 100 mV ± 1.5% ± 5dgts ± 2.0% ± 5dgts 800 volts AC
600V 1V ± 1.5% ± 5dgts ± 2.0% ± 5dgts 800 volts AC
1Although 400mV AC can be selected by the RANGE key, the Prova 2000A is not designed to measure AC millivolts. Consequently no accuracy is listed for the 400mV AC range. For accurate millivolt measurements, see: Multimeters
Resistance: (Autorange & Manual. Open Circuit Voltage ? 0.5V)
Range Resolution Accuracy Overload Protection
400? 0.1? ± 1.5% ± 3dgts 600 volts AC
4k? 1? ± 1.5% ± 3dgts 600 volts AC
40k? 10? ± 1.5% ± 3dgts 600 volts AC
400k? 100? ± 1.5% ± 3dgts 600 volts AC
4M? 1k? ± 1.5% ± 3dgts 600 volts AC
40M? 10k? ± 1.5% ± 3dgts 600 volts AC
Continuity: (Overload protection 600VAC. Open circuit voltage 0.4V)
Range Resolution Accuracy Beeping
40?-400? 0.1? ± 1.0% ± 2dgts < 40.0?
Diode Test:
Range Resolution Accuracy Overload Protection
3V 0.001V ± 2.5% ± 5dgts 600 volts AC
Frequency: (OL protection 600VAC. Sensitivity selected with RANGE button)
Range Resolution Accuracy Sensitivity
100Hz-1000kHz 0.01Hz 100Hz ± 0.5% ± 2dgts 10mV/100mV/1V
General specifications:
Applicable standards EN 61010-2-032, CAT III 600V, Pollution Degree 2
Conductor size Cable diameter 55mm approximately (2.2?)
Bus Bar 65mm(D) x 24mm(W), (2.6? x 0.9?)
Display 3¾ digit LCD with 40 segment bar graph
Operating temperature 4°C to 50°C
Dimensions 271(L) x 112(W) x 46(H)mm, (10.7? x 4.4? x 1.8?)
Weight 647g (batteries included), (1½ lb approx.)
Accessories included Users Manual, 9V battery, Carrying Bag